Home > Prodotti > Strumenti da processo > Regolatori > Panel-mounted pH Analog Indicator with Self Diagnostic Test – HI8510

SKU: HI8510 Categorie: ,

Panel-mounted pH Analog Indicator with Self Diagnostic Test – HI8510

589,00 + 22% IVA

9946 disponibili (ordinabile)

Guadagna 589 punti
Richiedi informazioni

HI8510 is ideal for monitoring pH in process control. It provides highly accurate pH measurements and displays values on an easy-to-read LCD. BNC input, amplified probe input and input from transmitter are supported.

  • Automatic temperature compensation ensures fast, accurately compensated readings
  • Displays values on the easy to read LCD
  • A removable, transparent splash-proof cover protects the front panel
  • ± 5V power output and input terminals for amplified electrodes

Prodotti indispensabili da aggiungere al tuo carrello

Tempi di spedizione
Non a stock, spedizione prevista entro 10 giorni lavorativi
Garanzia diretta dal produttore
Assistenza clienti sempre a disposizione
Maggiori informazioni
Specifiche tecniche
Recensioni
Downloads
Info Ordine

Easy Installation

HI8510 is provided with membrane keypads on the front panel, large display, and auto-diagnostic functions to check pH electrode and instrument status.

Easy pH Electrode Connection

Allows for quick and easy connection to any pH meter or transmitter.

Self-testing Diagnostics

Built-in autodiagnostic functions to enable the user to check and troubleshoot any malfunctions. The functions are made via front panel keys to isolate the cause of malfunction whether it is due to pH electrode contamination, internal offset circuit or the amplifier circuit.

Protective Cover

A removable, transparent splash-proof cover protects the front panel.

Recorder Terminals

To enhance troubleshooting and the ability to record data while monitoring, simply attach a recording device to the instrument’s 4 to 20 mA output contacts, conveniently located on the front panel, to obtain a copy of the results on demand.

Automatic Temperature Compensation

A 2-wire Pt100 can be connected to provide automatic temperature compensation of measurements.

Scala pH: 0.00 to 14.00 pH
Risoluzione pH: 0.01 pH
Accuratezza pH: ±0.02 pH (0 to 100 °C)/ ±0.05 pH (-20 to 0 °C) (electrode); ±0.5% (transmitter)
Calibrazione pH: 0.01 pH
Compensazione della Temperatura pH: 0.00 to 14.00 pH
Segnale Uscita Analogica: ±0.02 pH (0 to 100 °C)/ ±0.05 pH (-20 to 0 °C) (electrode); ±0.5% (transmitter)
Retroilluminazione: Continuous ON
Voltaggio di Uscita: ±5 Vcc; 150 mA max load for amplified electrodes
Componenti Esterne Incluse: Black anodized aluminium body; front and back with ABS; transparent splash-proof front cover
Alimentatore: 115 or 230 Vac; 60/50 Hz
Condizioni di Utilizzo: -10 to 50°C (14 to 122°F); RH max 95% non-condensing
Dimensioni: 141 x 69 mm (5.6 x 2.7'')
Peso: 1.00 kg (2.2 lb.)